PERMANENT CAUTERY HOOK INSTRUMMENT
Report
- Report Number
- 2955842-2012-01489
- Event Type
- Malfunction
- Date Received
- December 31, 2012
- Date of Event
- November 28, 2012
- Report Date
- November 28, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE INSTRUMENT WAS RETURNED WITH THE CONDUCTOR CAP DETACHED FROM THE DISTAL END. THE CAP IS INTACT, BUT HAS A CRACKED SECTION AROUND THE OUTER EDGE. THE CONDUCTOR WIRE IS EXPOSED AT THE DISTAL END AS A RESULT OF CAP DISLODGMENT. THE YAW PULLEY BOSS FEATURE THAT MATES WITH CAP IS BROKEN OFF. ENGINEERING EVALUATION ALSO FOUND THAT THE CLEVIS HAS AN AXIALLY ALIGNED CRACK HALFWAY BETWEEN THE EARS. THE LOCATION OF THE CRACK IS ON THE SAME SIDE AS THE CONDUCTOR WIRE/CAP, WHICH INDICATES OVERLOADING AT THE TIP OCCURRED WITH WHEN THE WRIST WAS PITCHED. ENGINEERING CONCLUDED THAT THE DAMAGE WAS LIKELY DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.
IT WAS REPORTED THAT THE ROUND PLASTIC PIECE FROM THE JOINT ON THE PERMANENT CAUTERY HOOK INSTRUMENT CAME OUT. LATE SUBMISSION OF THIS MEDWATCH REPORT IS DUE TO A REPORTING OVERSIGHT BY THE RESPONSIBLE COMPLAINT HANDLER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERMANENT CAUTERY HOOK INSTRUMMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420183-04 | M10100210 944 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SURGICAL SYSTEM, ACCESSORIES AND ESU. |