FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2891770 · Received December 31, 2012

Report

Report Number
3004209178-2012-91639
Event Type
Injury
Date Received
December 31, 2012
Date of Event
December 7, 2012
Report Date
December 11, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING INCLUDING THE DISPLACEMENT TEST. THE DEVICE HAD CRACKED BATTERY TUBE THREADS. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING IN THE EMERGENCY ROOM DUE TO DIABETES KETOACIDOSIS AND HIGH BLOOD GLUCOSE OF 582MG/DL. THE CUSTOMER EXPERIENCED VOMITING PRIOR TO HIS ADMISSION. TROUBLESHOOTING WAS DECLINED AS THE DOCTOR RECOMMENDED THE DEVICE BE REPLACED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-522NAS

Patients

Seq Age Sex Outcome Treatment
1 34 YR Hospitalization