FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2891592 · Received December 7, 2012

Report

Report Number
1720753-2012-10222
Event Type
Malfunction
Date Received
December 7, 2012
Date of Event
November 22, 2012
Report Date
December 7, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE HIGH VOLTAGE SUPPLY REGULATOR, FILAMENT DRIVE, AND GENERATOR DRIVER WERE REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A GENERATOR ERROR MESSAGE. THIS ERROR MESSAGE APPEARS WHEN THE SYSTEM DETECTS AN ERROR IN ANY OF THE REGISTERS ASSOCIATED WITH THE HIGH VOLTAGE GENERATOR. THE SYSTEM SHUTS DOWN WHEN THIS OCCURS. THERE IS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1