FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2891441 · Received December 31, 2012

Report

Report Number
2050012-2012-01983
Event Type
Malfunction
Date Received
December 31, 2012
Date of Event
December 6, 2012
Report Date
December 6, 2012
Manufacturer
BECKMAN COULTER INC.
Product Code
JGS
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

QC PRIOR TO AND AFTER THE EVENT WAS WITHIN LAB-ESTABLISHED RANGES. A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND REPLACED THE SODIUM (NA) ELECTRODE AND CARBON BRIDGE, WHICH RESOLVED THE ISSUE. INSTRUMENT PERFORMANCE WAS VERIFIED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) IN REGARDS TO OBTAINING FALSE LOW SODIUM (NA) RESULTS FOR SIX (6) PATIENT SAMPLES GENERATED ON THE UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER. THE FALSE RESULTS WERE REPORTED OUT OF THE LABORATORY. WHEN THE ISSUE WAS DISCOVERED, THE SAMPLES WERE RERUN ON ANOTHER INSTRUMENT IN THE LABORATORY AND REPORTS WERE AMENDED. THE RESULTS PROVIDED BY THE CUSTOMER ARE SHOWN IN THIS REPORT. THERE WAS NO IMPACT TO PATIENT OR PATIENT TREATMENT AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JGS BECKMAN COULTER INC. DXC 600 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1 46 YR