FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER
MDR report key: 2891441
·
Received December 31, 2012
Report
- Report Number
- 2050012-2012-01983
- Event Type
- Malfunction
- Date Received
- December 31, 2012
- Date of Event
- December 6, 2012
- Report Date
- December 6, 2012
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JGS
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
QC PRIOR TO AND AFTER THE EVENT WAS WITHIN LAB-ESTABLISHED RANGES. A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND REPLACED THE SODIUM (NA) ELECTRODE AND CARBON BRIDGE, WHICH RESOLVED THE ISSUE. INSTRUMENT PERFORMANCE WAS VERIFIED.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) IN REGARDS TO OBTAINING FALSE LOW SODIUM (NA) RESULTS FOR SIX (6) PATIENT SAMPLES GENERATED ON THE UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER. THE FALSE RESULTS WERE REPORTED OUT OF THE LABORATORY. WHEN THE ISSUE WAS DISCOVERED, THE SAMPLES WERE RERUN ON ANOTHER INSTRUMENT IN THE LABORATORY AND REPORTS WERE AMENDED. THE RESULTS PROVIDED BY THE CUSTOMER ARE SHOWN IN THIS REPORT. THERE WAS NO IMPACT TO PATIENT OR PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JGS | BECKMAN COULTER INC. | DXC 600 PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR |