PCA3 MEDNET WIRELESS
Report
- Report Number
- 9615050-2012-01577
- Event Type
- Malfunction
- Date Received
- December 21, 2012
- Date of Event
- November 30, 2012
- Report Date
- November 30, 2012
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- MEA
- PMA / PMN Number
- K042800
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
TESTING AND INVESTIGATION FOUND THE DEVICE DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION. THIS WAS DUE TO A PIN THAT WAS LIFTED FROM THE CONDUCTOR TRACE ON THE BOTTOM OF THE AUDIO TRANSDUCER WHICH DISABLED THE OUTPUT. THE CAUSE OF THE LIFTED PIN FROM THE CONDUCTOR TRACE ON THE AUDIO TRANSDUCER COULD NOT BE DETERMINED. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THE DEVICE DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION. THE DEVICE WAS RETURNED TO THE BIOMEDICAL ENGINEERING DEPARTMENT FOR AN UNSPECIFIED REASON. NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PCA3 MEDNET WIRELESS | 80MEA | MEA | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |