FDA Adverse Event
Malfunction
Summary report: N
MICRO DRILL MEDIUM STRAIGHT ATTACHMENT
MDR report key: 2891104
·
Received November 14, 2012
Report
- Report Number
- 1811755-2012-04052
- Event Type
- Malfunction
- Date Received
- November 14, 2012
- Date of Event
- October 12, 2012
- Report Date
- October 12, 2012
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER'S COMPLAINT WAS CONFIRMED; AN UNK BUR WAS STUCK IN THE ATTACHMENT. THE BURR DID NOT HAVE A J NOTCH WHICH IS ACCEPTED BY THIS ATTACHMENT AND RECOMMENDED IN THE IFU. THE STRYKER IFU ALSO RECOMMENDS THE USE OF ONLY STRYKER APPROVED COMPONENTS AND ACCESSORIES AS THE ATTACHMENT MAY CAUSE A NON-STRYKER BUR TO BREAK AND IF THE WRONG TYPE OF BUR IS USED.
Description of Event or Problem · 1
A MICRO DRILL MEDIUM STRAIGHT ATTACHMENT WAS SENT FOR SERVICE BECAUSE SOMETHING WAS STUCK IN IT. DURING EVAL OF THE DEVICE, AN UNK BROKEN BURR WAS NOTED IN THE ATTACHMENT. NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICRO DRILL MEDIUM STRAIGHT ATTACHMENT | ERL | STRYKER INSTRUMENTS KALAMAZOO | 11131 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |