FDA Adverse Event Malfunction Summary report: N

SYSTEM 6 ASEPTIC HOUSING ASSY

MDR report key: 2891099 · Received November 14, 2012

Report

Report Number
1811755-2012-04056
Event Type
Malfunction
Date Received
November 14, 2012
Date of Event
October 12, 2012
Report Date
October 12, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
KIJ
PMA / PMN Number
K972367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

AN INITIAL CORRECTIVE ACTION HAS TAKEN PLACE. THIS ISSUE HAS OCCURRED PREDOMINATELY IN (B)(6). AN EXTENSIVE INVESTIGATION CONDUCTED BY STRYKER INSTRUMENTS HAS DETERMINED THAT THE ROOT CAUSE OF THIS ISSUE IS RELATED TO THE RIGOROUS REPROCESSING METHODS USED IN (B)(6). THE (B)(6) GOVERNMENT SENDS OUT A CIRCULAR THAT RECOMMENDS THE HOUSINGS BE STERILIZED FOR 18 MINUTES AT 134 DEGREES C. MANUFACTURER INVESTIGATIVE TESTING INDICATES THE DEVICES IN THESE COMPLAINTS WERE NOT LASTING THE EXPECTED LIFE DUE TO THE COMBINATION OF CHEMICALS AND TIME EXPOSED TO HIGH HEAT. THE IFU FOR THIS PRODUCT RECOMMENDS 4 MINUTES EXPOSURE TIME AT 132-134 DEGREES C.; THEREFORE, THE REPROCESSING METHODS USED IN (B)(6) ARE NOT IN LINE WITH THE IFU. THE ROOT CAUSE FOR THESE TYPES OF CUSTOMER COMPLAINTS IS RELATED TO USER ERROR (I.E. FAILURE TO FOLLOW INSTRUCTIONS CONTAINED WITHIN THE IFU). STRYKER INSTRUMENTS HAS NOT RECEIVED ANY COMPLAINTS OF THIS NATURE IN WHICH IT IS ALLEGED THAT A SERIOUS INJURY HAS OCCURRED. THE PROBABILITY OF ACTUAL SERIOUS INJURY IS REMOTE BASED ON ACTUAL COMPLAINT DATA AND ESTIMATED DEVICE USAGE. THE SITUATION CONTINUES TO BE MONITORED AND REMAINS UNDER INVESTIGATION WHILE ONGOING REGULAR DISCUSSIONS ARE OCCURRING WITH (B)(6).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DELRIN BALL IN THE LOCKING MECHANISM OF THE ASEPTIC BATTERY HOUSING WAS MISSING. NOTHING FELL INTO A SURGICAL SITE, AS THIS WAS DETERMINED WHILE INSPECTING THE DEVICE. THERE WAS NO PT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYSTEM 6 ASEPTIC HOUSING ASSY KIJ STRYKER INSTRUMENTS KALAMAZOO 11186

Patients

Seq Age Sex Outcome Treatment
1 UNK