FDA Adverse Event Summary report: N

TRILOGY 100

MDR report key: 2890932 · Received December 26, 2012

Report

Report Number
2518422-2012-02505
Date Received
December 26, 2012
Date of Event
December 11, 2012
Report Date
December 11, 2012
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A FAILURE RELATED TO THE REMOTE ALARM CONNECTOR WAS OBSERVED. THE DEVICE'S INTERFACE BOARD WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE EVALUATION. THE DEVICE WAS NOT IN PT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC. 1054097 1054097

Patients

Seq Age Sex Outcome Treatment
1