FDA Adverse Event Summary report: N

BRILLANCE ICT SP

MDR report key: 2890920 · Received December 21, 2012

Report

Report Number
1525965-2012-00125
Date Received
December 21, 2012
Report Date
December 7, 2012
Manufacturer
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
Product Code
JAK
PMA / PMN Number
K060937
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NOTE: WE HAVE NOT COMPLETED OUR INVESTIGATION OF THIS EVENT. WE WILL FILE A FOLLOW-UP MDR AT THE COMPLETION OF THE INVESTIGATION. (B)(4).

Description of Event or Problem · 1

PHILIPS RECEIVED A REPORT THAT THE CUSTOMER REPORTED WHILE PERFORMING A PRESCAN FOR A RENAL BIOPSY, THE OPERATOR AND PT DETECTED A BURNING SMELL AND SMOKE COMING OUT OF TOP OF GANTRY. THE SMOKE DETECTOR WENT OFF AND THE PT WAS REMOVED FROM THE TABLE AND THE AREA WAS LOCKED DOWN AND FIRE DEPARTMENT WAS CALLED. THE FIRE DEPARTMENT EVALUATED THE ISSUE AND THERE WAS NO FIRE, ONLY SMOKE. THE PHILIPS FIELD SERVICE ENGINEER (FSE) WAS CALLED TO THE SITE AND EVALUATED THE SYSTEM AND DETERMINED THAT THE ISSUE WAS DUE TO A FAILED UPPER LINEAR INDUCTION MOTOR (LIM MOTOR).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BRILLANCE ICT SP JAK PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. 728311