FDA Adverse Event Injury Summary report: N

OCTRODE

MDR report key: 2890910 · Received December 21, 2012

Report

Report Number
1627487-2012-14255
Event Type
Injury
Date Received
December 21, 2012
Date of Event
November 21, 2012
Report Date
December 3, 2012
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

METHOD: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: REVIEW OF THE DHR FOUND A NONCONFORMANCE RELATED TO THE PRODUCT LOT. HOWEVER, THE INDIVIDUAL AFFECTED DEVICE WAS REWORKED AND ALL DEVICES WITHIN THE LOT MET ACCEPTANCE CRITERIA. THEREFORE, THE DHR ANOMALY IS NOT RELATED TO THE ALLEGED DEVICE COMPLAINT. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2012-14254. IT WAS REPORTED THE PT HAD AN INFECTION WHEN HIS TRIAL LEADS WERE REMOVED. THE PHYSICIAN NOTED A YELLOWISH DRAINAGE AND THE AREA WAS TENDER TO THE TOUCH. THE PT WAS TREATED WITH ANTIBIOTICS AND THE INFECTION HAS CLEARED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OCTRODE SCS LEAD LGW ST. JUDE MEDICAL - NEUROMODULATION 3086 3760279

Patients

Seq Age Sex Outcome Treatment
1 63 YR Required Intervention