FDA Adverse Event Malfunction Summary report: N

C-FLEX ASPHERIC

MDR report key: 2890572 · Received November 26, 2012

Report

Report Number
9611165-2012-00076
Event Type
Malfunction
Date Received
November 26, 2012
Date of Event
October 24, 2012
Report Date
November 1, 2012
Manufacturer
RAYNER INTRAOCULAR LENSES LTD.
Product Code
HQL
PMA / PMN Number
P060011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE REFERENCE (B)(4) HAS BEEN ALLOCATED TO THIS CASE BY RAYNER INTRAOCULAR LENSES LTD. RAYNER INTRAOCULAR LENSES LTD HAVE BEEN ADVISED BY THE (B)(4) DISTRIBUTOR THAT THE PHYSICIAN EXPLANTED THE C - FLEX ASPHERIC 970C INTRAOCULAR LENS AND IMPLANTED A BACK-UP INTRAOCULAR LENS OF THE SAME MODEL AND POWER IN THE ORIGINAL PLANNED SURGERY SESSION WITHOUT ANY ADVERSE EVENT TO THE PT. THE CONDITION OF THE PT POST-OPERATIVELY HAS BEEN DESCRIBED AS "EXCELLENT" BY THE HEALTHCARE FACILITY. THE EVENT DESCRIPTION PROVIDED BY THE CANADIAN DISTRIBUTOR INDICATES THAT THE EVENT OCCURRED AS A RESULT OF USER ERROR, THEREFORE ADDITIONAL LENS LOADING TRAINING, SINCE THIS EVENT, HAS BEEN PROVIDED TO THE PHYSICIAN TO PREVENT THE RECURRENCE OF THIS ISSUE. OUR REVIEW OF PRODUCTION RECORDS FOR THE C-FLEX ASPHERIC 970C BATCH 121E31751 SHOWED THAT ALL MFG AND QUALITY CHECKS WERE CONDUCTED WITH SUCCESSFUL RESULTS. ALL LENSES RELEASED FOR DISTRIBUTION FROM THIS BATCH WERE WITHIN TOLERANCE, MET SPECIFICATION LIMITS AND WERE WITHOUT DEFECTS. OUR EXISTING VIGILANCE DATA SINCE THE MONTH OF MANUFACTURE OF THE C-FLEX ASPHERIC 970C INTRAOCULAR LENS (DECEMBER 2011) WAS REVIEWED IN ORDER TO DETERMINE IF ANY TRENDS EXISTED. THIS REVIEW CONCLUDED THAT NO OTHER INCIDENTS, OF ANY TYPE, HAVE BEEN RECEIVING AGAINST THE C-FLEX ASPHERIC 970C BATCH 121E31751.

Description of Event or Problem · 1

RAYNER INTRAOCULAR LENSES LIMITED RECEIVED NOTIFICATION FROM THE (B)(4) DISTRIBUTOR OF AN EVENT THAT OCCURRED DURING USE OF A C-FLEX ASPHERIC 970C INTRAOCULAR LENS. THE EVENT DESCRIPTION PROVIDED STATES "HAPTIC BROKE WHEN INJECTED."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 C-FLEX ASPHERIC HQL - INTRAOCULAR LENS HQL RAYNER INTRAOCULAR LENSES LTD. 970C 121E31751

Patients

Seq Age Sex Outcome Treatment
1