FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2890553 · Received December 31, 2012

Report

Report Number
2531779-2012-15245
Event Type
Malfunction
Date Received
December 31, 2012
Report Date
December 5, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE INSULIN PUMP HAS BEEN RETURNED AND ADDITIONAL INVESTIGATION WAS PERFORMED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: REVIEW OF THE DOWNLOAD HISTORY DID NOT REVEAL ANY LARGE PRIME AMOUNTS RECORDED. ON TESTING, THE PUMP POWERED ON NORMALLY. EVALUATION REVEALED A DISCOLORED DISPLAY SCREEN, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION. THE OWNER'S BOOKLET INSTRUCTS THE USER TO CALL ANIMAS CUSTOMER SERVICE IF THE USER SUSPECTS THAT THE PRODUCT IS DAMAGED. A REWIND, LOAD AND PRIME SEQUENCE WAS PERFORMED SUCCESSFULLY. THE FORCE SENSOR CALIBRATION TEST REVEALED THE FORCE SENSOR TO BE OUT OF SPECIFICATION. THE PUMP WAS OPENED FOR INVESTIGATION AND REVEALED CONTAMINATION ON THE FORCE SENSOR ASSEMBLY.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED CONTAMINATION OF THE FORCE SENSOR ASSEMBLY AND THE FORCE SENSOR OUT OF SPECIFICATION. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2012.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 30 YR