FDA Adverse Event Malfunction Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 2890216 · Received December 28, 2012

Report

Report Number
2017865-2012-10733
Event Type
Malfunction
Date Received
December 28, 2012
Date of Event
November 15, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0457
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4).INTERNAL INSULATION ABRASION WAS NOTED AT 8.8-10.3CM FROM THE ELECTRODE TIP. THE ETFE COATING WAS INTACT AT THIS LOCATION.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ASYMPTOMATIC PATIENT PRESENTED IN CLINIC FOR FOLLOW UP. FLUOROSCOPY REVEALED EXTERNALIZED CONDUCTORS. NO ELECTRICAL ANOMALIES WERE DETECTED. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1590/65 NA

Patients

Seq Age Sex Outcome Treatment
1 67 YR