FDA Adverse Event Malfunction Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 2890214 · Received December 28, 2012

Report

Report Number
2017865-2012-10725
Event Type
Malfunction
Date Received
December 28, 2012
Date of Event
November 21, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0457
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A FRACTURE OF THE RV CABLES WERE FOUND AT 9.2CM FROM THE DISTAL TIP CONSISTENT WITH FATIGUE.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD WAS EXPLANTED DUE TO INSULATION ANOMALY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1582/65 NA

Patients

Seq Age Sex Outcome Treatment
1