FDA Adverse Event Injury Summary report: N

ATTACHABLE VOIDING STOOL

MDR report key: 28902 · Received August 30, 1995

Report

Report Number
28902
Event Type
Injury
Date Received
August 30, 1995
Date of Event
February 3, 1995
Report Date
February 20, 1995
Manufacturer
LIEBEL-FLARSHEIM CO.
Product Code
EYH
Product Problem
Yes
Report Source
User Facility report
Reporter Location
LA, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

CYSTO TABLE BEING USED WITH "VOIDING STOOL" ATTACHED. PT IN SITTING POSITION ON VOIDING STOOL. WHILE THE PHYSICIAN DOING THE "FILLING AND VOIDING CYSTOMETROGRAM" (FOR PT WITH STRESS INCONTINENCE AND VAGINAL VAULT PROLAPSE) THE TABLE WAS BEING POSITIONED , LOUD NOISE WAS HEARD, AND PT SEEN AT AN ANGLE. THE BED WAS TILTED SO PT'S FALL WAS BROKEN. PT DID NOT FALL. IT APPEARED THAT SCREWS ON VOIDING STOOL GOT CAUGHT ON SIDE OF CYSTOL TABLE, CAUSING STOOL TO BREAK OFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTACHABLE VOIDING STOOL ATTACHABLE VOIDING STOOL EYH LIEBEL-FLARSHEIM CO.

Patients

Seq Age Sex Outcome Treatment
1 74 YR Required Intervention