FDA Adverse Event Malfunction Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 2890176 · Received December 28, 2012

Report

Report Number
2017865-2012-10648
Event Type
Malfunction
Date Received
December 28, 2012
Date of Event
October 17, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0457
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED EXTERNALIZED CONDUCTORS DUE TO INTERNAL INSULATION ABRASION WERE FOUND AT 7.6CM-9.7CM FROM THE DISTAL TIP. THE SILICONE INSULATION WAS ABRADED AND THE RV AND SENSING CONDUCTORS WERE VISIBLE. EXTERNAL INSULATION ABRASION WAS FOUND AT 40.3 CM-40.7CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS INTACT AT THESE LOCATIONS. (B)(6. NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.