FDA Adverse Event
Malfunction
Summary report: N
RIATA ACTIVE FIXATION
MDR report key: 2890167
·
Received December 28, 2012
Report
- Report Number
- 2017865-2012-10719
- Event Type
- Malfunction
- Date Received
- December 28, 2012
- Date of Event
- November 26, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- Z0457
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A PARTIAL LEAD MEASURING 50.0CM OF THE DISTAL TIP ELECTRODE WAS RETURNED. EXTERNALIZED CONDUCTORS DUE TO INTERNAL INSULATION ABRASION WERE FOUND AT 8.8CM TO 11.6CM FROM THE DISTAL TIP. THE SILICONE INSULATION WAS ABRADED AND THE RV CONDUCTOR WAS VISIBLE. THE ETFE COATING WAS ABRADED AT THIS LOCATION.
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT PRESENTED IN THE HOSPITAL FOR A NORMAL FOLLOW UP. EXTERNALIZED CONDUCTORS WERE OBSERVED VIA FLUOROSCOPY. THE LEAD WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RIATA ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1582/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |