FDA Adverse Event
Malfunction
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
MDR report key: 2890083
·
Received December 28, 2012
Report
- Report Number
- 2017865-2012-10782
- Event Type
- Malfunction
- Date Received
- December 28, 2012
- Date of Event
- October 18, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
INSULATION DEGRADATION WAS NOTED AT 13.3-14.1CM FROM THE DISTAL TIP. THE SILICONE INSULATION WAS INTACT AT THE SAME LOCATION.
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.
Description of Event or Problem · 1
IT WAS REPORTED THAT OVERSENSING DUE TO NOISE AND HIGH IMPEDANCE WERE OBSERVED. LEAD WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7121Q/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |