FDA Adverse Event Malfunction Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 2890083 · Received December 28, 2012

Report

Report Number
2017865-2012-10782
Event Type
Malfunction
Date Received
December 28, 2012
Date of Event
October 18, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INSULATION DEGRADATION WAS NOTED AT 13.3-14.1CM FROM THE DISTAL TIP. THE SILICONE INSULATION WAS INTACT AT THE SAME LOCATION.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.

Description of Event or Problem · 1

IT WAS REPORTED THAT OVERSENSING DUE TO NOISE AND HIGH IMPEDANCE WERE OBSERVED. LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7121Q/58 NA

Patients

Seq Age Sex Outcome Treatment
1 52 YR