FDA Adverse Event Injury Summary report: N

UTILITY PLIER (WEINGART STYLE)

MDR report key: 2887392 · Received December 19, 2012

Report

Report Number
2086211-2012-00001
Event Type
Injury
Date Received
December 19, 2012
Date of Event
November 19, 2012
Report Date
November 26, 2012
Manufacturer
DENTSPLY GAC
Product Code
JEX
PMA / PMN Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

BECAUSE EVALUATION OF THE UNIT INVOLVED IS NOT COMPLETE AS OF THIS REPORT AND SINCE THIS ISSUE COULD NECESSITATE MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION, IT MUST BE PRESUMED THAT THE DEVICE MALFUNCTIONED AND THAT THE MALFUNCTION WOULD BE LIKELY TO CAUSE/CONTRIBUTE TO A SERIOUS INJURY SHOULD IT RECUR. AS SUCH, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE DEVICE WAS FOUND TO BE WITHIN SPECIFICATION. DURING THE INVESTIGATION IT WAS FOUND THAT THE DOCTOR WAS USING UTILITY PLIERS AS A CRIMPABLE STOP, WHICH IS NOT THE INTENDED USE OF THIS DEVICE, AND LIKELY THE CAUSE FOR THE BEAK BREAKAGE.

Description of Event or Problem · 1

IN THIS EVENT IT WAS REPORTED THAT THE TIP OF A UTILITY PLIER (WEINGART STYLE) SEPARATED INSIDE OF A PATIENT'S MOUTH AND CHIPPED THEIR TOOTH. AS A RESULT A COMPOSITE CROWN WAS PLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UTILITY PLIER (WEINGART STYLE) JEX DENTSPLY GAC 07-06-1

Patients

Seq Age Sex Outcome Treatment
1 9 YR Required Intervention CRIMPABLE STOP