FDA Adverse Event Malfunction Summary report: N

HEAD RING DRIVE (BRW & CRW)

MDR report key: 2885671 · Received December 19, 2012

Report

Report Number
1222895-2012-00053
Event Type
Malfunction
Date Received
December 19, 2012
Report Date
December 19, 2012
Manufacturer
INTEGRA BURLINGTON, MA, INC.
Product Code
HAW
PMA / PMN Number
K811452
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

TO DATE, THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVAL. AN INVESTIGATION HAS BEEN INITIATED BASED UPON THE REPORTED INFO.

Description of Event or Problem · 1

HRD CRW HEAD RING DRIVE WELD ON NOTCHED RIM OF THE STAINLESS POSTHOLDER SEPARATED FROM THE DISC. IT WAS RETAINED ONLY BY TWO STAINLESS DOWELS - IF THESE ALSO FAILED THE POST WOULD HAVE ROTATED ON THE POSTHOLDER, ENDANGERING THE PT. THE FAILURE WAS DISCOVERED DURING PRE-SURGERY INSPECTION OF THE HEADRING, AND THE POSTHOLDER WAS REPLACED WITH A SPARE, WITHOUT FURTHER PROBLEMS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEAD RING DRIVE (BRW & CRW) NA HAW INTEGRA BURLINGTON, MA, INC.

Patients

Seq Age Sex Outcome Treatment
1