FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 2885342 · Received November 20, 2012

Report

Report Number
1314492-2012-00440
Event Type
Malfunction
Date Received
November 20, 2012
Report Date
September 28, 2012
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SIGMA DEVICE EVALUATION FOUND THE #7, #8 AND #9 KEYS TO BE INOPERABLE. IT WAS OBSERVED THAT WHEN ANY REMAINING FUNCTIONAL KEYS ARE SELECTED, AN AUTOMATIC OUTPUT OF THE #7 KEY WILL OCCUR FOLLOWING THE SELECTED KEY'S FUNCTION (E.G. WHEN THE #1 KEY IS PRESSED 17 WILL BE DISPLAYED. WHEN IN ALPHABETIC MODE AND THE ABC KEY IS SELECTED, AS WILL BE DISPLAYED INTERFERING WITH THE MDL DRUG SEARCH). SIGMA'S ENGINEERING INVESTIGATION IS IN PROGRESS. WHEN COMPLETE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DATE OF EVENT IS UNK.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE #7 KEY ON A PUMP KEYPAD IS "STUCK."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1