FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2885295 · Received November 21, 2012

Report

Report Number
3008642652-2012-03066
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
October 31, 2012
Report Date
November 19, 2012
Manufacturer
ZOLL CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WILL NOT POWER ON) HAS BEEN CONFIRMED. AS RECEIVED, THE MONITOR WOULD NOT POWER ON. THE CAUSE OF THE MONITOR NOT POWERING ON WAS A FLASH MEMORY FAILURE (COMPONENTS U102 AND U105) ON THE C/A BOARD SN (B)(4). THE FLASH MEMORY HAD AN INTERMITTENT CONNECTION, WHICH WAS DISCOVERED THROUGH THE USE OF A TARGET MEMORY TEST. DURING THE FLASH MEMORY TEST THE MONITOR PRODUCED A SEGMENTATION FAULT. THE INTERMITTENT CONNECTION IS LIKELY DUE TO A FRACTURED SOLDER CONNECTION INDUCED BY MECHANICAL STRESS. THE EXACT SOURCE OF THE MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED BUT THE FRACTURED CONNECTION MAY HAVE BEEN ACCELERATED THROUGH ROUGH HANDLING. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE COMPONENTS. THE PATIENT RECEIVED A REPLACEMENT MONITOR.

Description of Event or Problem · 1

A (B)(6) MALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS MONITOR WOULD NOT POWER ON. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 53 YR