FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 2885293
·
Received November 21, 2012
Report
- Report Number
- 3008642652-2012-03090
- Event Type
- Malfunction
- Date Received
- November 21, 2012
- Date of Event
- November 1, 2012
- Report Date
- November 21, 2012
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (ADJUST BELT OR CHECK BELT MESSAGES) WAS CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT FAILED A FALL-OFF TEST. THE CAUSE FOR THE TEST FAILURE AND THE REPORTED ALARMS WAS ISOLATED TO AN OPEN WHITE WIRE (DRIVEN GROUND) IN THE ELECTRODE BELT TRUNK CABLE. THE ROOT CAUSE FOR THE OPEN WIRE COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE OPEN DRIVEN GROUND WIRE. THE PATIENT RECEIVED A REPLACEMENT BATTERY ELECTRODE BELT.
Description of Event or Problem · 1
(B)(6) MALE PATIENT'S NURSE CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HE WAS RECEIVING ADJUST BELT OR CHECK BELT MESSAGES. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |