FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2885293 · Received November 21, 2012

Report

Report Number
3008642652-2012-03090
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
November 1, 2012
Report Date
November 21, 2012
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (ADJUST BELT OR CHECK BELT MESSAGES) WAS CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT FAILED A FALL-OFF TEST. THE CAUSE FOR THE TEST FAILURE AND THE REPORTED ALARMS WAS ISOLATED TO AN OPEN WHITE WIRE (DRIVEN GROUND) IN THE ELECTRODE BELT TRUNK CABLE. THE ROOT CAUSE FOR THE OPEN WIRE COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE OPEN DRIVEN GROUND WIRE. THE PATIENT RECEIVED A REPLACEMENT BATTERY ELECTRODE BELT.

Description of Event or Problem · 1

(B)(6) MALE PATIENT'S NURSE CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HE WAS RECEIVING ADJUST BELT OR CHECK BELT MESSAGES. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 58 YR