FDA Adverse Event Injury Summary report: N

3M BAIR HUGGER WARMING UNIT

MDR report key: 2885090 · Received December 12, 2012

Report

Report Number
2110898-2012-00065
Event Type
Injury
Date Received
December 12, 2012
Date of Event
November 11, 2012
Report Date
November 14, 2012
Manufacturer
3M HEALTH CARE
Product Code
DWJ
PMA / PMN Number
960167
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

3M WILL NEED TO EVALUATE THE UNIT TO CONFIRM IF BIOMED RESULTS ARE ACCURATE. THIS IS A VERY OLD UNIT. THEY STATED THE UNIT WAS ON THE HIGHEST TEMPERATURE WHEN BEING USED. THE WARMING BLANKET USED WITH THE WARMING UNIT WAS: 3M BAIR HUGGER FULL BODY BLANKET, MODEL 300, ORDER NUMBER 30000, NO LOT NUMBER WAS PROVIDED. PMA/510(K): APPLICABLE TO BOTH WARMING UNIT AND BLANKET. THE MODEL 500R HAS AN INDEPENDENT OVER-TEMPERATURE ALARM. THERE IS AN OVER-TEMPERATURE INDICATOR ON THE USER INTERFACE AND IF THERE IS AN OVER-TEMPERATURE CONDITION, THE INDICATOR ILLUMINATES AND AN AUDIBLE ALARM SOUNDS AND THE HEATER SHUTS OFF.

Description of Event or Problem · 1

CUSTOMER REPORTED A HYPOTHERMIC, VERY CRITICAL, (B)(6) MALE PT WITH MULTIPLE ISSUES, SUSTAINED 2ND TO 3RD DEGREE BURNS WITH BLISTERS TO 8-10% OF HIS LEFT LATERAL LEG FROM THE KNEE TO ANKLE FOLLOWING 3.5 HOURS OF WARMING WITH A 500/OPERATING ROOM WARMING UNIT. REPORTEDLY, PT WAS SENT TO WOUND CARE WITH THE NO INFO ABOUT TREATMENT. A PT BLANKET WAS PLACED OVER THE WARMING BLANKET (MODEL 300) AND A SHEET WAS PLACED BETWEEN THE PT AND THE WARMING BLANKET AS IS THEIR PRACTICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 3M BAIR HUGGER WARMING UNIT SYSTEM, THERMAL REGULATING 870.5900 DWJ DWJ 3M HEALTH CARE 500OR NA

Patients

Seq Age Sex Outcome Treatment
1 49 YR Other NONE SPECIFIED