FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2884959 · Received December 4, 2012

Report

Report Number
1720753-2012-10040
Event Type
Malfunction
Date Received
December 4, 2012
Date of Event
November 15, 2012
Report Date
December 4, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE FILAMENT DRIVER BOARD WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A CHARGER FAILURE ERROR MESSAGE WHICH PREVENTED THE SYSTEM FROM BOOTING UP. THERE IS NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1