FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2884853
·
Received December 3, 2012
Report
- Report Number
- 1720753-2012-09900
- Event Type
- Malfunction
- Date Received
- December 3, 2012
- Date of Event
- November 19, 2012
- Report Date
- November 30, 2012
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE REP EVALUATED THE SYSTEM AND RELOADED SOFTWARE AND REPLACED THE UPS. THE SYSTEM OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A FILE CORRUPTION ERROR. THIS ERROR CAUSES THE SYSTEM TO LOCK UP OR NOT TO BOOT. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |