FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2884847 · Received December 3, 2012

Report

Report Number
1720753-2012-09906
Event Type
Malfunction
Date Received
December 3, 2012
Date of Event
November 14, 2012
Report Date
November 30, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE X-RAY TUBE WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM WAS ARCING AND DISPLAYED OVERLOAD FAULTS DURING PROCEDURES AND CONCLUDED THAT THE X-RAY TUBE NEEDED REPLACEMENT. THIS ERROR MAY CAUSE THE SYSTEM TO SHUT DOWN. THERE IS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1