FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2884823 · Received December 3, 2012

Report

Report Number
1720753-2012-09929
Event Type
Malfunction
Date Received
December 3, 2012
Date of Event
November 21, 2012
Report Date
November 30, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE RANDOM ACCESS MEMORY AND CPU BOARDS WERE RESEATED. THE SYS WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYS WOULD NOT BOOT UP. THERE IS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1