FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK SPIRIT

MDR report key: 2884783 · Received December 3, 2012

Report

Report Number
2183996-2012-01864
Event Type
Malfunction
Date Received
December 3, 2012
Date of Event
July 23, 2012
Report Date
April 18, 2013
Manufacturer
ROCHE INSULIN DELIVERY SYSTEMS INC.
Product Code
LZG
PMA / PMN Number
K060876
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT WAS RECEIVED FOR EVALUATION ON (B)(4) 2013. THE COMPLAINT CANNOT BE VERIFIED, AND THE PRODUCT MEETS THE SPECIFICATIONS REGARDING THE CUSTOMERS ALLEGATION. THE DELIVERY ACCURACY AND THE OCCLUSION LIMITS WERE TESTED WITHIN THE PUMP DRIVE TEST ON THE DIAGNOSTIC TEST SYSTEM AND BOTH MEET THE SPECIFICATIONS. ALL ALARM FUNCTIONS WERE TESTED SUCCESSFULLY AND NO MALFUNCTION COULD BE OBSERVED DURING THE TECHNICAL INVESTIGATION. THE BATTERY COVER PASSED THE OPTICAL INSPECTION.

Description of Event or Problem · 1

ON (B)(6) 2012 THE PT REPORTED THAT THE INFUSION DEVICE CHANGES FROM RUN TO STOP MODE W/O WARNING. THIS HAS BEEN OCCURRING FOR 3.5-4 MONTHS. SHE STATED THE ISSUE LAST OCCURRED ABOUT A MONTH AGO. THE PT REVIEWED THE ALARM HISTORY AND AN E4 (OCCLUSION) ERROR WAS LISTED ON (B)(6) 2012. AT THAT TIME THE PT GAVE THE INFUSION DEVICE TO A NURSE WHO ASSISTS HER AND THE NURSE WAS ABLE TO CLEAR ERROR. SHE STATED THAT SHE IS CERTAIN THE INFUSION DEVICE DID NOT DISPLAY AN ALERT OR ERROR MESSAGE WHEN IT WENT INTO STOP MODE. THE PT DID NOT REQUIRE TREATMENT WHEN IT WENT INTO STOP MODE. THE PT DID NOT REQUIRE TREATMENT FROM A HEALTH CARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. THE INFUSION DEVICE WS REPLACED. THE INFUSION DEVICE, BATTERY, AND BATTERY COVER WAS REQUESTED TO BE RETURNED FOR EVAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCU-CHEK SPIRIT LZG ROCHE INSULIN DELIVERY SYSTEMS INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 49 YR RELATED ACCESSORIES| INSULIN INFUSION PUMP| INSULIN