FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 2884295 · Received December 22, 2012

Report

Report Number
3008382007-2012-08849
Event Type
Malfunction
Date Received
December 22, 2012
Report Date
December 8, 2012
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE METER WAS FOUND TO HAVE AN OPEN/SHORTED CAPACITOR ISSUE. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2012, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING ERROR 1. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3377812

Patients

Seq Age Sex Outcome Treatment
1 72 YR