FDA Adverse Event Malfunction Summary report: N

V60 VENTILATOR

MDR report key: 2884243 · Received December 21, 2012

Report

Report Number
2031642-2012-00559
Event Type
Malfunction
Date Received
December 21, 2012
Report Date
December 14, 2012
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
MNT
PMA / PMN Number
K082660
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

FACTORY ANALYSIS OF A RETURNED POWER MANAGEMENT PCB BOARD REVEALED A COMPONENT FAILURE THAT MAY RESULT IN THE UNIT SHUTTING DOWN DURING USE IN NORMAL VENTILATION MODE OPERATION. UPON LOSS OF POWER, A POWER FAIL ALARM WILL ACTIVATE AND IF IN USE ALTERNATIVE VENTILATION SHOULD BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 V60 VENTILATOR VENTILATOR, CONTINUOUS MNT RESPIRONICS CALIFORNIA, INC. V60 NA

Patients

Seq Age Sex Outcome Treatment
1