FDA Adverse Event Injury Summary report: N

8MM MCS TIP COVER ACCESSORY

MDR report key: 2884107 · Received December 21, 2012

Report

Report Number
2955842-2012-01412
Event Type
Injury
Date Received
December 21, 2012
Report Date
August 2, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ISI HAS CONTACTED THE INITIAL REPORTER CONCERNING THE REPORTED EVENT. A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS PROVIDED.

Description of Event or Problem · 1

ON (B)(6) 2012, DR (B)(6) AT (B)(6) HOSPITAL FORWARDED TO INTUITIVE SURGICAL AN ARTICLE, TITLED ROBOTIC INSTRUMENT INSULATION FAILURE: INITIAL REPORT OF A POTENTIAL SOURCE OF PATIENT INJURY PUBLISHED BY DR. ADAM MUES, GEOFFREY BOX AND RONNEY ABAZA. THE ARTICLE ALLEGES THAT FROM JULY 2008 - JANUARY 2009 12 MCS TIP COVER ACCESSORIES, INSTALLED ON THE MONOPOLAR CURVED SCISSORS INSTRUMENTS FAILED INTRA-OPERATIVELY DURING DA VINCI ROBOTIC PROCEDURES AND WITH 3 OF THE TIP COVER FAILURES RESULTING IN PATIENT INJURIES. THE FOLLOWING DETAILS CONCERNING THE SECOND PATIENT INJURY AS INDICATED IN THE ARTICLE IS AS FOLLOWS: DURING A DA VINCI UROLOGICAL PROCEDURE, WHILE PERFORMING A RETROPERITONEAL LYMPHADENECTOMY USING THE MONOPOLAR CURVED SCISSORS INSTRUMENT WITH THE MCS TIP COVER INSTALLED, THE PATIENT'S SMALL BOWEL WAS INJURED. DAMAGE TO AFFECTED AREA WAS RECOGNIZED IMMEDIATELY AND WAS REPAIRED BY REAPPROXIMATION OF THE SEROSA USING A SILK SUTURE. INSPECTION OF THE TIP COVER ACCESSORY BY THE SITE FOUND THAT THE TIP COVER EXHIBITED A HOLE AT THE JUNCTION OF THE SILICON AND GRAY PLASTIC SHAFT ON THE TIP COVER. PER THE INFORMATION INDICATED IN THE REPORT, THE TIP COVER ACCESSORY WAS INSPECTED PRIOR TO USE, WAS INSTALLED ON THE MCS INSTRUMENT USING THE TIP COVER INSTALLATION TOOL WITH NO OBVIOUS MECHANICAL IMPERFECTIONS OBSERVED IN THE TIP COVER. NO OTHER DETAILS CONCERNING THE PATIENT WERE PROVIDED. MEDWATCH REPORT 2955842-2008-01353 WAS SUBMITTED TO THE FDA CONCERNING THE FIRST PATIENT INJURY AND MEDWATCH REPORT 2955842-2012-01414 WAS SUBMITTED FOR THE THIRD REPORT OF A PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8MM MCS TIP COVER ACCESSORY ENDOSCOPIC ELECTROSURGICAL INSTRUMENT ACCESSORY NAY INTUITIVE SURGICAL,INC. 400180

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention DAVINCI S SURG SYSTEM, INSTRUMENT AND ACCESSORIES