FDA Adverse Event Malfunction Summary report: N

5MM MONOPOLAR CAUTERY INSTRUMENT

MDR report key: 2884087 · Received December 21, 2012

Report

Report Number
2955842-2012-00820
Event Type
Malfunction
Date Received
December 21, 2012
Date of Event
October 8, 2012
Report Date
October 8, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. COMPLAINT OF CAN SEE METAL BEYOND THE INSULATION IS CONFIRMED. BLACK TUBE INSULATION IS DAMAGED AT THE DISTAL END. INSULATION IS GOUGED ON ONE SIDE AND HAS A .100 X .040 PIECE MISSING ROUGHLY .240 ABOVE THE SNAKE WRIST. INSULATION EXHIBITS VARIOUS OTHER SCRATCHES/GOUGES AT THE DISTAL END WITHOUT MATERIAL REMOVAL. ADDITIONAL OBSERVATION NOT REPORTED BY SITE IS A BENT MAIN TUBE. MAIN TUBE IS BENT ROUGHLY 4 BELOW THE DISTAL TIP. BEND ANGLE IS ROUGHLY 5 DEGREES, BUT BEND BECOMES MORE APPARENT WHEN TUBE IS ROTATED ALONG ROLL AXIS. EVIDENCE NOT CONCLUSIVE, BUT BENDING DAMAGE IS LIKELY DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.

Description of Event or Problem · 1

IT WAS OBSERVED DURING CENTRAL PROCESSING THAT THE 5MM MONOPOLAR CAUTERY INSTRUMENT HAD A HOLE IN THE INSULATION. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 5MM MONOPOLAR CAUTERY INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420142-04 S10120501 147

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES