FDA Adverse Event Malfunction Summary report: N

VITROS 5600 INTEGRATED SYSTEM

MDR report key: 2883694 · Received December 21, 2012

Report

Report Number
1319681-2012-00257
Event Type
Malfunction
Date Received
December 21, 2012
Date of Event
November 30, 2012
Report Date
December 21, 2012
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
Product Code
JJE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION DETERMINED THAT AN UNEXPECTED DELAY IN OBTAINING INTRA-OPERATIVE VITROS IPTH RESULT WAS OBSERVED FOR SINGLE PATIENT SAMPLE ON TWO DIFFERENT VITROS 5600 SYSTEMS. THE INVESTIGATION DETERMINED THAT THE DELAY WAS CAUSED DUE TO TEMPORARY MECHANICAL ISSUES ON BOTH VITROS 5600 SYSTEMS. BOTH ISSUES WERE TEMPORARY AND WERE RESOLVED BY ROUTINE OPERATOR ACTIONS ALLOWING THE SAMPLE TO BE PROCESSED SUCCESSFULLY. HOWEVER, THERE WAS NO MALFUNCTION OF THE VITROS 5600 SYSTEM OR VITROS IPTH REAGENT IDENTIFIED. THE ROOT CAUSE OF THIS EVENT IS INSTRUMENT RELATED.

Description of Event or Problem · 1

THE CUSTOMER OBSERVED AN UNEXPECTED DELAY IN OBTAINING INTRA-OPERATIVE VITROS IPTH RESULT FOR A SINGLE PATIENT SAMPLE ON TWO DIFFERENT VITROS 5600 SYSTEMS. A DELAY OF APPROXIMATELY ONE HOUR WAS OBSERVED WITH RESPECT TO THE EXPECTED TURNAROUND TIME. DUE TO THIS DELAY IN OBTAINING VITROS IPTH RESULTS, THE PATIENT WAS HELD UNDER ANESTHESIA FOR AN ADDITIONAL 30 MINUTES WHILE UNDERGOING A SURGICAL PROCEDURE. THE LIKELIHOOD OF AN INCREMENTAL INCREASE IN THE INHERENT HEALTH RISKS TO THE PATIENT DUE TO EXTENDED TIME UNDER ANESTHESIA IS NOT REMOTE. ONCE THE VITROS IPTH RESULTS WERE OBTAINED, THE SURGICAL PROCEDURE WAS COMPLETED. THERE WAS NO REPORT OF HARM TO THE PATIENT AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER TWO OF TWO MDR'S FOR THIS EVENT. TWO 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS TWO DEVICES WERE INVOLVED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITROS 5600 INTEGRATED SYSTEM CHEMISTRY ANALYZER JJE ORTHO-CLINICAL DIAGNOSTICS

Patients

Seq Age Sex Outcome Treatment
1