ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2012-14989
- Event Type
- Malfunction
- Date Received
- December 21, 2012
- Report Date
- November 30, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2012 WITH THE FOLLOWING FINDINGS: THERE IS NO PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE KEYPAD WAS INTACT, THERE WAS NO DAMAGE OR PEELING OBSERVED. DURING TESTING THE CONTRAST BUTTON WAS FOUND TO BE INTERMITTENTLY UNRESPONSIVE, WHICH HAS NO EFFECT ON INSULIN DELIVERY; ALL OTHER BUTTONS RESPONDED APPROPRIATELY. DURING INVESTIGATION, EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL KEYPAD CONTACTS. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A DIM/FADED DISPLAY SCREEN, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION.
ON (B)(6) 2012, THE PATIENT CONTACTED ANIMAS STATING THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE AND STICKING. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |