VA LOCKSCR Ø2.4 SELF-TAP L18 TAN
Report
- Report Number
- 1719045-2012-01433
- Event Type
- Malfunction
- Date Received
- December 21, 2012
- Report Date
- November 22, 2012
- Manufacturer
- SYNTHES MONUMENT
- Product Code
- ODX
- PMA / PMN Number
- K103242
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. THE DEVICE WAS RECEIVED, THE INVESTIGATION COULD NOT BE COMPLETED, AND NO CONCLUSION COULD BE DRAWN, AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM.
THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE SCREW RECEIVED IS CONTAINED IN A PROPERLY SEALED SYNTHES PACKAGE WITH NO EVIDENCE OF TEARS OR OPENINGS OF ANY TYPE. THE PRINTED LABEL INDICATES THAT IT CONTAINS A 04.210.118, LOT NUMBER 6971717. UPON EXAMINATION THROUGH THE PACKAGING WITH A 10X MAGNIFIER, CHIP WRAP WAS FOUND AT THE NECK OF THE SCREW. THE PACKAGE WAS OPENED TO ALLOW FURTHER EXAMINATION. NO OTHER BURRS/CHIPS WERE NOTED. DHR WAS REVIEWED WITH NO COMPLAINT RELATED ANOMALIES NOTED.
A DEVICE REPORT FROM (B)(6) INDICATED A HOSPITAL IN (B)(6) REPORTED: SCREWS IN UNOPENED PACKAGES WERE TAKEN FROM STOCK AND IT WAS NOTICED THERE WERE VISIBLE CHIPS ON ALL SCREW HEADS. THERE WAS NO PATIENT INVOLVEMENT. THIS IS 6 OF 6 REPORTS FOR THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VA LOCKSCR Ø2.4 SELF-TAP L18 TAN | VA LOCKSCREW | ODX | SYNTHES MONUMENT | 6971717 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |