FDA Adverse Event
Malfunction
Summary report: N
TERUMO ADVANCED PERFUSION SYSTEM 1
MDR report key: 2883086
·
Received December 18, 2012
Report
- Report Number
- 1828100-2012-01605
- Event Type
- Malfunction
- Date Received
- December 18, 2012
- Date of Event
- December 5, 2012
- Report Date
- December 5, 2012
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DTQ
- PMA / PMN Number
- K022947
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE REPORTED COMPLAINT WAS CONFIRMED BY THE FIELD SERVICE REP (FSR). THE BATTERIES AND BATTERY CABLE WERE REPLACED AND THE UNIT OPERATED TO CLINICAL USE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
Description of Event or Problem · 1
THE FIELD SERVICE REPRESENTATIVE (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, HE FOUND CORROSION ON ONE WIRE TERMINAL OF THE BATTERY CABLE. THE FSR SCRAPED OFF CORROSION ON THE CABLE. THERE WAS NO CORROSION ON THE BATTERY ITSELF. SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TERUMO ADVANCED PERFUSION SYSTEM 1 | DTQ | TERUMO CARDIOVASCULAR SYSTEMS CORP. | 801763 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |