FDA Adverse Event Malfunction Summary report: N

TERUMO ADVANCED PERFUSION SYSTEM 1

MDR report key: 2883086 · Received December 18, 2012

Report

Report Number
1828100-2012-01605
Event Type
Malfunction
Date Received
December 18, 2012
Date of Event
December 5, 2012
Report Date
December 5, 2012
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DTQ
PMA / PMN Number
K022947
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED COMPLAINT WAS CONFIRMED BY THE FIELD SERVICE REP (FSR). THE BATTERIES AND BATTERY CABLE WERE REPLACED AND THE UNIT OPERATED TO CLINICAL USE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

THE FIELD SERVICE REPRESENTATIVE (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, HE FOUND CORROSION ON ONE WIRE TERMINAL OF THE BATTERY CABLE. THE FSR SCRAPED OFF CORROSION ON THE CABLE. THERE WAS NO CORROSION ON THE BATTERY ITSELF. SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TERUMO ADVANCED PERFUSION SYSTEM 1 DTQ TERUMO CARDIOVASCULAR SYSTEMS CORP. 801763

Patients

Seq Age Sex Outcome Treatment
1