FDA Adverse Event Injury Summary report: N

OT ULTRALINK METER

MDR report key: 2882930 · Received December 21, 2012

Report

Report Number
3008382007-2012-08788
Event Type
Injury
Date Received
December 21, 2012
Date of Event
November 12, 2012
Report Date
December 14, 2012
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K073231
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

FOLLOW-UP (2/22/2013)-DEVICE EVALUATION: THE LAY USER/PATIENTS PRODUCT(S) HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT WERE FOUND HAVE RESULTS OUTSIDE OF THE CONTROL SOLUTION RANGE. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED. THE RETAIN TEST STRIPS WERE TESTED AND THEY PASSED TESTING WITH NO FAULTS FOUND. A DHR (DEVICE HISTORY RECORD) WAS COMPLETED FOR THIS PRODUCT AND NO DEVIATIONS, NON-CONFORMANCES, OR REWORKS WERE OBSERVED.

Description of Event or Problem · 1

ON (B)(6) 2012, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING HIS ONETOUCH ULTRALINK METER READ INACCURATELY HIGH COMPARED TO HIS FEELINGS/ NORMAL BLOOD GLUCOSE RESULTS. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE ALLEGED ISSUE BEGAN IN (B)(6) 2012. THE PATIENT OBTAINED A BLOOD GLUCOSE RESULT OF "AROUND 300 MG/DL" WITH THE SUBJECT METER. IT IS NOT CLEAR HOW THE PATIENT MANAGES HIS DIABETES; HOWEVER, THAT MORNING, THE PATIENT ADMINISTERED SELF 17 UNITS HUMULIN INSULIN. AN HOUR AFTER THE START OF THE ALLEGED ISSUE, THE PATIENT CLAIMS HE FELT SYMPTOMS OF CLAMMY AND WEAKNESS. TREATMENT WAS NOT SPECIFIED. ON THE MORNING OF (B)(6) 2012, THE PATIENT REPORTEDLY VISITED HIS PHYSICIAN'S OFFICE AND OBTAINED A BLOOD GLUCOSE RESULT OF "AROUND 125 MG/DL" WITH THE PHYSICIAN'S METER. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THE SUBJECT METER WAS SET TO THE CORRECT UNIT OF MEASUREMENT. THE PATIENT DID NOT HAVE CONTROL SOLUTION TO PERFORM QUALITY CONTROL TESTING. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE, THE PATIENT CLAIMS HE OBTAINED AN INACCURATE HIGH READING ON THE SUBJECT METER, ADMINISTERED TREATMENT BASED ON THE ALLEGED RESULT, AND REPORTEDLY DEVELOPED SYMPTOMS SUGGESTIVE OF SERIOUS INJURY AFTER THE ALLEGED METER ISSUE BEGAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT ULTRALINK METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3373538

Patients

Seq Age Sex Outcome Treatment
1 62 YR Life Threatening