FDA Adverse Event Malfunction Summary report: N

ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM

MDR report key: 2882823 · Received December 20, 2012

Report

Report Number
2024168-2012-08242
Event Type
Malfunction
Date Received
December 20, 2012
Date of Event
November 28, 2012
Report Date
November 28, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
NIO
PMA / PMN Number
P110028
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE BROKEN (OPEN) HANDLE WAS ABLE TO BE CONFIRMED. BASED ON A VISUAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING DEVICE PREPARATION AND PRIOR TO USE, THE ABSOLUTE PRO STENT DELIVERY SYSTEM (SDS) WAS FLUSHED AND WHEN THE SYRINGE WAS UNSCREWED FROM THE FLUSH PORT THE HANDLE GAPPED OPEN AND SEPARATED. THERE WAS NO PATIENT INVOLVEMENT. THE DEVICE WAS NOT USED. A DIFFERENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM SELF EXPANDING STENT SYSTEM NIO AV-TEMECULA-CT 2072061

Patients

Seq Age Sex Outcome Treatment
1