FDA Adverse Event Malfunction Summary report: N

AC-POWERED ADJUSTABLE HOSPITAL BED, INCLUDING RAILS

MDR report key: 2882750 · Received December 20, 2012

Report

Report Number
3003433498-2012-00115
Event Type
Malfunction
Date Received
December 20, 2012
Report Date
December 19, 2012
Manufacturer
CARROLL HEALTHCARE
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

A REPORT HAS BEEN RECEIVED THAT OCCURRED AT A CANADIAN LONG TERM CARE USER FACILITY. IT WAS LEARNED BY THE DEALER THAT THE T CABLE ONLY WORKS WHEN IT IS WIGGLED TO GET IT TO WORK. THE PART WAS REPLACED UNDER WARRANTY. NO INJURY TO THE INVOLVED END USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AC-POWERED ADJUSTABLE HOSPITAL BED, INCLUDING RAILS 880.5100 FNL CARROLL HEALTHCARE IHCS7

Patients

Seq Age Sex Outcome Treatment
1 Other