FDA Adverse Event Malfunction Summary report: N

WAVESCAN WAVEFRONT SYSTEM WS1

MDR report key: 2882459 · Received December 20, 2012

Report

Report Number
3006695864-2012-00191
Event Type
Malfunction
Date Received
December 20, 2012
Date of Event
August 14, 2012
Report Date
December 18, 2012
Manufacturer
ABBOTT MEDICAL OPTICS
Product Code
HKO
PMA / PMN Number
K000327
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

THE CAUSE OF THE COMPUTER'S POWER SUPPLY COMPONENTS FAILING AND EMITTING SMOKE WAS NOT DETERMINED. THE COMPUTER WAS REPAIRED WITH A NEW POWER SUPPLY INSTALLED AND WAS TESTED. THE COMPUTER OPERATED CORRECTLY FOLLOWING REPAIR. ALL PERTINENT INFORMATION AVAILABLE TO AMO HAS BEEN SUBMITTED.

Description of Event or Problem · 1

THE AMO FIELD SERVICE ENGINEER RETURNED THE CUSTOMER'S WAVESCAN COMPUTER TO AMO FOR EVALUATION INTO THE COMPUTER NOT FUNCTIONING. SERVICE DEPOT EVALUATED THE COMPUTER AND WHEN THE POWER SUPPLY WAS POWERED UP, SMOKE WAS OBSERVED BEING EMITTED FROM THE POWER SUPPLY AND THE COMPUTER WAS NON FUNCTIONAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WAVESCAN WAVEFRONT SYSTEM WS1 REFRACTIVE MEASUREMENT HKO ABBOTT MEDICAL OPTICS WS1

Patients

Seq Age Sex Outcome Treatment
1