PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2882444 · Received December 20, 2012

Report

Report Number
2024168-2012-08214
Event Type
Injury
Date Received
December 20, 2012
Date of Event
December 10, 2012
Report Date
December 10, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FROM THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. THREE UNUSED STERILE PROGLIDE DEVICES WITH THE SAME LOT NUMBER, 20906J1, AS THE COMPLAINT DEVICE WERE RETURNED FOR EVALUATION. FUNCTIONAL TESTING WAS PERFORMED AND THE DEVICES PASSED WITH ACCEPTABLE RESULTS. NO MANUFACTURING OR ABNORMAL OBSERVATIONS WERE DETECTED. A CAUSE FOR THE COMPLAINT DEVICE REPORTED EXPERIENCE COULD NOT BE DETERMINED BASED ON THE INVESTIGATION FINDINGS FROM THE TESTED SAMPLES. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS PROVIDED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE SECOND PROGLIDE DEVICE IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF A COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE AFTER A RIGHT CORONARY INTERVENTIONAL PROCEDURE, USING A 6F SHEATH. REPORTEDLY, THE LOOP WAS OUTSIDE OF THE SHEATH AND THE KNOT LOOKED STRANGE. THE KNOT PULLED OUT OF THE PATIENT'S ANATOMY WHEN TENSION WAS APPLIED TO ADVANCE THE KNOT. A SECOND PROGLIDE DEVICE WAS USED AND A CUFF MISS OCCURRED. A NON-ABBOTT DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. THE TECHNICIAN IS REPORTEDLY TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 20906J1

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SHEATH: 6FANGIOMAX