ACCESS
Report
- Report Number
- 1416980-2012-07843
- Event Type
- Malfunction
- Date Received
- December 20, 2012
- Date of Event
- December 4, 2012
- Report Date
- December 4, 2012
- Manufacturer
- BAXTER HEALTHCARE - CALI
- Product Code
- FPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- PHARMACIST
Narratives
(B)(4). ADDITIONAL INFORMATION: THE SAMPLE WAS RECEIVED FOR EVALUATION AND INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. A BATCH REVIEW CANNOT BE PERFORMED SINCE THERE WAS NO LOT NUMBER PROVIDED. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP WILL BE SUBMITTED.
(B)(4). EVALUATION SUMMARY: ONE ACTUAL SAMPLE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED AND NO DEFECTS WERE OBSERVED. AN UNDER WATER PRESSURE TEST AT 8 PSI WAS PERFORMED. THERE WAS AN ABSENCE OF SOLVENT BETWEEN THE TUBING AND INJECTION SITE AND A LEAK WAS OBSERVED. THE REPORTED CONDITION WAS CONFIRMED. THE ROOT CAUSE WAS NOT DETERMINED. ADDITIONAL INFORMATION: THIS ISSUE IS BEING INVESTIGATED THROUGH A CAPA. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
(B)(4). THE SAMPLE IS REPORTED TO BE AVAILABLE FOR EVALUATION. IF THE SAMPLE IS RECEIVED OR ADDITIONAL INFORMATION BECOMES AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. THIS IS A PRODUCT NOT DISTRIBUTED IN THE US AND DOES NOT HAVE A 510K NUMBER.
THE CUSTOMER REPORTED TO BAXTER (B)(4) OF A SOLUTION SET IN WHICH A LEAK IS OBSERVED AT THE MOLDED Y. THE PROCESS STEP IS NOT SPECIFIED. THERE IS NO REPORT OF PATIENT INVOLVEMENT, INJURY/ADVERSE EVENTS, OR MEDICAL INTERVENTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS | SET, ADMINISTRATION, INTRAVASCULAR | FPA | BAXTER HEALTHCARE - CALI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |