FDA Adverse Event
Malfunction
Summary report: N
LIGACLIP** ENDOSCOPIC CLIP APPLIER
MDR report key: 2882315
·
Received December 20, 2012
Report
- Report Number
- 3005075853-2012-05851
- Event Type
- Malfunction
- Date Received
- December 20, 2012
- Date of Event
- December 11, 2012
- Report Date
- December 14, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE ANALYSIS RESULTS OF THE DEVICE FOUND THAT IT WAS RECEIVED IN GOOD VISUAL CONDITION. THE JAWS WERE FOUND PROPERLY ALIGNED. IN AN ATTEMPT TO REPLICATE THE INCIDENT REPORTED THE DEVICE WAS FUNCTIONALLY EVALUATED. UPON FIRING OF THE DEVICE, THE REMAINING CLIPS WERE EJECTED AND THEN, IT LOCKED OUT AS INTENDED. IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSE THE FOUND CONDITION. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A LAPAROSCOPIC ASSISTED DISTAL GASTRECTOMY PROCEDURE, TWO CLIPS WERE FIRED AT A TIME. BESIDES, SCISSORING OCCURRED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGACLIP** ENDOSCOPIC CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |