VITROS IMMUNODIAGNOSTIC PRODUCTS TROPONIN I ES REAGENT PACK
Report
- Report Number
- 3007111389-2012-00256
- Event Type
- Malfunction
- Date Received
- December 20, 2012
- Date of Event
- December 3, 2012
- Report Date
- December 20, 2012
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- MMI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION DETERMINED THAT A NON-REPRODUCIBLE, HIGHER THAN EXPECTED VITROS TROP I ES RESULT WAS OBTAINED FROM A SINGLE PATIENT RUN ON THE VITROS 5600 SYSTEM. THERE WAS NO EVIDENCE TO SUGGEST AN INSTRUMENT/ REAGENT MALFUNCTION HAD OCCURRED. ADDITIONALLY, THE SAMPLE WAS NOT PROCESSED IN ACCORDANCE WITH THE TUBE MANUFACTURER'S RECOMMENDATIONS. IT IS LIKELY THAT CELLULAR DEBRIS, DUE TO POOR SAMPLE PREPARATION, WAS PRESENT IN THE AFFECTED SAMPLE, ALTHOUGH THIS COULD NOT BE CONFIRMED. A DEFINITIVE ROOT CAUSE FOR THE EVENT COULD NOT BE DETERMINED. HOWEVER, A PRE-ANALYTICAL SAMPLE PROCESSING ISSUE COULD NOT BE RULED OUT AS A CONTRIBUTING FACTOR.
THE CUSTOMER OBTAINED A NON-REPRODUCIBLE, HIGHER THAN EXPECTED VITROS TROP I ES RESULT (0.148 VS. EXPECTED RESULTS <0.012 NG/ML) FROM A SINGLE PATIENT RUN ON THE VITROS 5600 SYSTEM. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION. HOWEVER, THE HIGHER THAN EXPECTED RESULT WAS NOT REPORTED FROM THE LABORATORY. THE CUSTOMER RETESTED THE SAMPLE IN DUPLICATES PER CUSTOMER POLICY PRIOR TO REPORTING. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITROS IMMUNODIAGNOSTIC PRODUCTS TROPONIN I ES REAGENT PACK | IN-VITRO DIAGNOSTIC | MMI | ORTHO-CLINICAL DIAGNOSTICS | 1000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |