COMFORTDRIVE 200XDR
Report
- Report Number
- 3003637274-2012-00065
- Event Type
- Malfunction
- Date Received
- December 17, 2012
- Date of Event
- November 12, 2012
- Report Date
- November 26, 2012
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EFB
- PMA / PMN Number
- K080677
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- DENTIST
Narratives
THE ANALYSIS OF THE HANDPIECE DID SHOW THAT THE HEAD HAD A DENT. THIS CAUSED THE BACK CAP BUTTON TO STICK IN, CAUSING UNUSUAL INNER FRICTION, CAUSING FINALLY THE HEAD TO HEAT UP. THE DENT INDICATES THAT THE HANDPIECE RECEIVED A STRONG HIT, E.G. A DROP DURING THE REPROCESSING. ALSO THE BEARINGS HAVE BEEN GRITTY. IT IS LIKELY THAT THIS IS THE RESULT OF A STRONGER WEAR PROCESS DUE TO THE HIGH TEMPERATURES. THE USER INSTRUCTION REQUIRES A VISUAL AND FUNCTIONAL TEST PRIOR TO EACH TREATMENT TO ENSURE PRODUCT IS WITHIN SPECIFICATION, WHICH WOULD HENCE BE MANDATORY. REPORTED DUE TO THE 2 YEAR PRESUMPTION RULE.
DURING A STANDARD DENTAL TREATMENT THE HANDPIECE GOT HOT AND BURNED THE PT ON CHEEK APPROXIMATELY TO THE SIZE OF THE HANDPIECE HEAD. PT WAS GIVEN SOME SAMPLES OF KENOLOG ORABASE OINTMENT. NO MEDICAL CARE WAS NECESSARY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COMFORTDRIVE 200XDR | DENTAL HANDPIECE | EFB | KAVO DENTAL GMBH | 200XDR |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR |