GENTLEPOWER LUX CONTRA ANGLE 25LPA
Report
- Report Number
- 3003637274-2012-00063
- Event Type
- Malfunction
- Date Received
- December 17, 2012
- Date of Event
- November 19, 2012
- Report Date
- November 21, 2012
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- DENTIST
Narratives
THE ANALYSIS OF THE HANDPIECE DID SHOW THAT THE BEARINGS HAVE BEEN GRITTY AND THE HEAD WAS DENTED CAUSING THE BACKCAP BUTTON TO STICK IN. THIS CAUSED UNUSUAL INNER FRICTION CAUSING THE HEAD TO HEAT UP. THE DENT INDICATED THAT THE HANDPIECE WAS DROPPED OR MISHANDLED, E.G. DURING REPROCESSING. THE CONDITION OF THE BEARING MIGHT BE A NORMAL WEAR ISSUE, BUT IT IS LIKELY THAT DUE TO THE DENT AND THE HIGH TEMPERATURE A STRONGER WEAR TOOK PLACE. THE USER INSTRUCTION REQUIRES A VISUAL AND FUNCTIONAL INSPECTION PRIOR TO EACH TREATMENT WHICH WOULD HENCE BE MANDATORY AND SHOW WHETHER PRODUCT IS RUNNING WITHIN SPECIFICATION. REPORTED DUE TO THE 2 YEAR PRESUMPTION RULE.
DURING A STANDARD DENTAL TREATMENT ON THE LOWER BICUSPID THE HANDPIECE HEATED UP AND BURNED THE PT ON THE CHEEK TO HALF OF THE SIZE OF A DIME. HE WAS GIVEN SAMPLES OF PERIDEX RINSE AND KENALOG ORABASE. NO MEDICAL CARE WAS NECESSARY. THE DENTAL OFFICE DOES NOT RECALL THE NAME OF THE PT, THEREFORE THE AGE AND THE DATE OF EVENT CAN BE SUPPLIED ONLY APPROXIMATELY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWER LUX CONTRA ANGLE 25LPA | DENTAL HANDPIECE | EFB | KAVO DENTAL GMBH | 25LPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |