FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2881938 · Received December 20, 2012

Report

Report Number
2531779-2012-14920
Event Type
Malfunction
Date Received
December 20, 2012
Report Date
November 26, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012, WITH THE FOLLOWING FINDINGS: THE FORCE SENSOR PINS WERE FOUND TO BE DAMAGED. THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. DURING EVALUATION THE PUMP WAS ABLE TO REWIND, LOAD AND PRIME SUCCESSFULLY.

Description of Event or Problem · 1

THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012, WITH THE FOLLOWING FINDINGS: THE FORCE SENSOR PINS WERE FOUND TO BE DAMAGED. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 42 YR