HIP HEAD 32 MM

MDR report key: 2881805 · Received December 20, 2012

Report

Report Number
1818910-2012-84012
Event Type
Injury
Date Received
December 20, 2012
Date of Event
October 24, 2012
Report Date
December 3, 2012
Manufacturer
DEPUY ORTHOPAEDICS
Product Code
KWA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

VISUAL EXAMINATION OF THE RETURNED DEVICES FINDS EVIDENCE TO SUPPORT DISASSOCIATION OF THE LINER AND CUP WHILE IMPLANTED. IN THIS CASE A PORTION OF THE RIM OF THE POLYETHYLENE LINER HAS FRACTURED. IT IS EVIDENT THAT THE EDGE OF THE LINER WAS LOADED BY THE FEMORAL HEAD AND PATIENT BODY WEIGHT WHILE IMPLANTED. THERE ARE MACHINING LINES YET VISIBLE WITHIN THE INNER RADIUS OF THE LINER WHICH WOULD NOT BE AS OBVIOUSLY PRESENT HAD THE FEMORAL HEAD BEEN MORE CENTRALLY LOADED. THE EDGE LOADING OF THE LINER HAS PLACED PRESSURES ON THE MATERIAL NEAR THE RIM THAT WERE NOT INTENDED LEADING TO A FRACTURE OF THE MATERIAL AND SUBSEQUENT DISASSOCIATION OF THE LINER FROM THE CUP. REVIEW OF PROVIDED PATIENT X-RAYS CONFIRMS THE FEMORAL HEAD APPEARS TO BE ECCENTRICALLY LOCATED WITHIN THE POLYETHYLENE INSERT OF THE ACETABULAR CUP SUGGESTING THE LINER HAS DISASSOCIATED FROM THE ACETABULAR SHELL. A REVIEW OF DEVICE HISTORY RECORDS FINDS NO RELATED MANUFACTURING DEVIATIONS OR ANOMALIES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A SEARCH OF THE COMPLAINTS DATABASES FINDS NO OTHER REPORTS AGAINST THE PRODUCT/LOT CODE COMBINATION. THE INVESTIGATION CAN DRAW NO CONCLUSIONS WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN INDICATED. DEPUY CONSIDERS THIS INVESTIGATION CLOSED. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION THAT CHANGES THIS CONCLUSION BE RECEIVED, THE INVESTIGATION WILL BE REOPENED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

REVISION BECAUSE OF LOOSENING OF THE INLAY

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HIP HEAD 32 MM DEPUY HIP IMPLANT KWA DEPUY ORTHOPAEDICS UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 50 YR Required Intervention