FDA Adverse Event Injury Summary report: N

QUATTRODE

MDR report key: 2881420 · Received December 17, 2012

Report

Report Number
1627487-2012-14217
Event Type
Injury
Date Received
December 17, 2012
Date of Event
September 1, 2011
Report Date
December 4, 2011
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
GZB
PMA / PMN Number
K072462
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

METHOD: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: REVIEW OF THE DHR FOUND A NONCONFORMANCE RELATED TO THE PRODUCT LOT. HOWEVER, THE INDIVIDUAL AFFECTED DEVICE WAS REWORKED AND ALL DEVICES WITHIN THE LOT MET ACCEPTANCE CRITERIA. THEREFORE, THE DHR ANOMALY IS NOT RELATED TO THE ALLEGED DEVICE COMPLAINT. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 3 OF 5. REFERENCE MFR REPORT: 1627487-2012-14215, 14216, 14218, 14219. THE PATIENT WAS IMPLANTED WITH TWO QUATTRODE LEADS (3146) WITH THE SAME LOT NUMBER. THE PATIENT WAS ALSO IMPLANTED WITH TWO DUAL EXTENSIONS (3342) WITH THE SAME LOT NUMBER. IT WAS REPORTED THE PATIENT'S ENTIRE SCS SYSTEM WAS EXPLANTED DUE TO AN INFECTION. THE PATIENT DEVELOPED A PIMPLE OVER ONE OF THE PERIPHERAL LEADS. THE PHYSICIAN BROKE THE PIMPLE, AND TOOK A CULTURE WHICH CAME BACK POSITIVE FOR AN INFECTION. THE SYSTEM WAS THEN EXPLANTED TO BE ON THE SAFE SIDE SOMETIME IN (B)(6) 2011 (EXACT DATE UNKNOWN). THE PATIENT HAD A NEW SCS SYSTEM IMPLANTED ON (B)(6) 2011.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUATTRODE SCS LEAD GZB ST. JUDE MEDICAL - NEUROMODULATION 3166 2822544

Patients

Seq Age Sex Outcome Treatment
1 40 YR Required Intervention